K-PATENTS - PR-23-W 化學腐蝕性溶液濃度計--大管路型


特性:

  • PR-23-W,專為化學侵蝕性溶劑及超純化學製程設計,適合大型管路(直徑50,80,100mm/2", 3", 4")的製程監測,使用特殊材質,耐強酸強鹼。
  • 精確度:+/- 0.0002 Refractive Index, 相當於 +/- 0.1 % by weight.。
  • 測量範圍:
    標準:R.I. 1.3100 ~ 1.5400 (spinel prism)
    選擇:R.I. 1.2600 ~ 1.5000 (sapphire prism)
  • 樣品中的顆粒、氣泡或顏色改變不會影響測值精確度。
  • 數位原理:校正後不因時間而偏移,電極校正完成後無需再次校正。
  • 接液部份:採用PTFETeflon® or Kynar® Sapphire (Al2O3), Kalrez® and Spinel (MgAl2O4)等材質,可耐機械及化學侵蝕。另外,選配PELDTeflon® prism seal也能使用於HF應用上。
  • 外部材質:採用不鏽鋼AISI 316L (電極),polyamid 6.6 (electrical connector),不受化學侵蝕氣體或溶液溢出影響。

應用:

  • CHEMICALLY Aggressive Solutions 化學腐蝕溶液
  • Ultra-Pure fine Chemical Processes in large-scale production 大產量化學純化製程
  • Large pipelines 大尺寸管徑等應用

 

 

■      詳細規格表,請參見中文目錄。

 

Refractive Index range, standard Full range, nD1.3200...1.5300 (corresponds to 0...100% b.w.), sapphire prism
Refractive Index range, option nD1.2600...1.4700, sapphire prism
Accuracy nD± 0.0002, typically corresponds to ± 0.1% b.w. Repeatability nD± 0.0001, typically corresponds to ±0.05% b.w.)
Speed of response 1s undamped, damping time selectable up to 5 min
Patented CORE-Optics No mechanical adjustments and digital measurement with 3648 pixel CCD element, sodium D-line light emitting diode (LED), built-in Pt-1000 temperature sensor
Temperature compensation Automatic, digital compensation
Instrument verification With standard nD liquids and the documented procedure (ISO 9000)
Ambient temperature Sensor: -20 °C...45 °C (-4 °F...113 °F); Indicating transmitter: 0 °C...45 °C (32 °F...122 °F)
Process temperature -20 °C...130 °C (-4 °F... 266 °F)
Sensor protection class IP67, Nema 4X
Approvals for hazardous locations Non-Sparking/Non-Incendive version: ATEX certified for Zone 2 area, Ex II 3 G, EEx nA IIC T4 Gc; FM certified for Class I, Div. 2, Groups A, B, C & D hazardous locations, T6; CSA certified for Class I, Div. 2, Groups A, B, C & D, T4; IECEx certified, Ex nA IIC T4 Gc Intrinsically safe version: ATEX certified for Zone 0 area, Ex II 1 G, EEx ia IIC T4 Ga; FM certified for Class I, Div. 1, Groups A, B, C & D, T4 (for U.S.A. and Canada); IECEx certified, Ex ia IIC T4 Ga
Approval for electrical safety cCSAus certified, Equipment Class I, Poll. Deg. 2 / Transmitter, Poll. Deg. 3 / Sensor, Install. Cat. II
SENSOR PR-23-W Process connection with ETFE (Ethylene tetrafluoroethylene) lined Saunders (diaphragm) valve body 2", 3" or 4"; valve body material Graphite cast iron
Saunders body process connection DIN-flange DN50 PN 16/DIN-flange DN80 PN 16/DIN-flange DN100 PN 16/ANSI-flange 4" 150 lbs/ANSI-flange 3" 150 lbs/ANSI-flange 2" 150 lbs / JIS-flange 10k 50A/ JIS-flange 10k 80A/ JIS-flange 10k 100A
Sensor wetted parts ETFE (Ethylene tetrafluoroethylene) lining, prism gasket Teflon® PTFE, prism sapphire, O-ring Kalrez, adaptor sapphire
Sensor weight with 2" Saunders body 15 kg (33 lbs)/3" 26 kg (57 lbs)/4" 33 kg (73 lbs)
Process pressure max. 10 bar (145 psi)
INDICATING TRANSMITTER DTR
Display 320x240 pixel graphical LCD with LED backlight, keypad with 8 membrane keys
Current output Two independent current outputs, 4-20 mA, max. load 1000 Ohm, galvanic isolation 1500 VDC or AC (peak), hold function during prism wash
Fieldbus and industrial Ethernet connectivity Through Fieldbus converter to Modbus/TCP, Modbus RTU and Ethernet/ IP networks
Power AC input 100-240 VAC/50-60 Hz, optional 24 VDC, 30 VA
Alarms/Wash relays Two built-in signal relays, max. 250 V/3 A
Sensor connectivity One or two sensors can be connected to one DTR. Sensors independent with own parameter sets and usable in different applications. Two current outputs configurable independently to indicate process concentration or temperature of either sensor.
Indicating transmitter weight 4.5 kg (10 lbs)
Transmitter protection class Polycarbonate enclosure IP66, Type 4X (indoor use); AISI 304 Stainless steel enclosure IP66 (Indoor use)
INTERCONNECTING CABLE IEC 61158-2 compliant two-wire cable
Interconnecting cable length Standard 10 m (33 ft), max. 200 m (660 ft). Cable fittings to Indicating transmitter: European cable glands M20x1.5 or US conduit hubs