Online - CT-6110-RS Free Chlorine/Ozone Transmitter


The new design of CT-6110-RS can connect the free Cl electrode/ozone electrode/pH electrode to achieve a powerful accurate measurement


Powerful measurement with dual parameters

The CT-6110-RS transmitter has a rugged IP 65 enclosure and can be wall, panel, or pipe mounted in almost all industrial installations. Compatible with residual chlorine electrodes or ozone electrodes, the CT-6110-RS can also connect with pH electrodes for compensation which provides more stable and accurate measurement. In addition to inheriting the graphic menu and easy operation advantages of our intelligent series, the transmitter also contains RS model features including RS485 interface, measurement curve display, and the event logbook function.


Applications:

• Recycle water treatment

• Drinking water disinfection

• Beverage bottle disinfection

• Swimming pool

• Industrial process

• Water treatment


Features:

• Real-time monitoring of pH

• pH compensation for residual chlorine measurements provide accurate measurement and control

• Measured values can be selected for figure or chart display

• Trace mode can display historically measured value curve from 3-minute to 4-week program

• Equipped with events logbook function and password protection

• Easy operation, dual channel, analog and Modbus output transmitter

 

Model CT-6110-RS

Measurement Parameter

Residual Chlorine / Ozone / pH / Temp
Measurement Range Residual Chlorine 0.00~20.00ppm (mg/L), ±0.5% (± 1 Digit)
Ozone 0.00~2.00ppm (mg/L), ±0.5% (± 1 Digit)
pH -2.00~16.00pH, ±0.01pH (± 1 Digit)
Temperature -30.0~130.0°C, ±0.2°C (± 1 Digit)
Display LCM (graphic menu display) with auto/manual illumination function, available for Text mode, Trace mode, and Chart mode
Text Mode Digit text reading, simultaneous displays residual chlorine, O3, pH, temperature reading
Trace Mode Set up from 3 minutes to 4 weeks duration of the residual chlorine or O3 value trend graph to master the process history as well as the display of real-time residual chlorine reading at the bottom.
Chart Mode 3 minutes real-time dynamic residual chlorine measurement graph to monitor recent development in control as well as the display of realtime residual chlorine reading at the bottom
Temperature compensation Automatic with NTC 30K/PT1K or Manual adjustment
pH compensation Automatic compensation / Manual adjustment / Off
Output Signal Two isolated DC 0/4~20 mA corresponding to • Residual chlorine or O3 • pH or temperature, max load 500Ω One RS-485 (MODBUS RTU or MODBUS ASCII) interface
Set point Contact 240 VAC, 0.5A max. (recommended)
Activate Two individual relays corresponding to residual chlorine/O3 or pH, Hi/Lo selectable, limited programmable
Logbook 50 event records
Power Supply 100 ~ 240 VAC ±10%, 7W Max, 50~60 Hz
Installation Wall or Pipe or Panel mounting
Ambient Temperature 0 ~ 50°C
Storage Temperature -20 ~ 70°C
Dimensions 96 mm x 96 mm x 132 mm (H x W x D)
Cut off dimensions 93 mm x 93 mm (H x W)
Weight Approx. 0.5 kg
Protection IP 65 (NEMA 4X)

 

Calibration Method

• Zero: Pure water or activated charcoal filtered water (residual chlorine/O3 free water) calibration

• Span: 1. Input sampling DPD measurement result for residual chlorine calibration 2. Use a spectrophotometer with ozone reagent to obtain a reference value, then compare with transmitter measurement

• Clean/Auto Zero: Relay contact, ON: 0~99min. 59 sec. / OFF:0~999 hours 59 minutes


Transmitter Dimensions


Flow-Through Chamber Illustration (Optional)