Features:
感測器
■90度角散射光學監測原理設計,符合U.S. EPA 180.1規範
■封閉式迴路,測值不受外界光源干擾
■1/2/3點自動校正,單點校正可根據實驗室或現場分析調整
■內建自動去除氣泡設計,可避免氣泡干擾造成誤差
Model |
TD-100 |
Measuring range |
0.000~100.00NTU |
Resolution |
0.001(0.000~9.999 NTU) / 0.01(10.00~10.00 NTU) / Auto |
Accuracy |
±2% of reading from 0~40 NTU ±5% of reading from 40~100 NTU |
Repeatability |
±1% of full scale |
Light source |
White light |
Sample inlet fitting |
1/4” quick fitting with 1.5M hose (provided) |
Sample drain fitting |
1/2” hose connector |
Sample flow required |
200~750 c.c./min. |
Sample temperature |
0 ~ 50°C |
Cable length |
3 m |
Dimensions |
600 x 180 x 226mm (H x W x D) |
Weight |
3.7 kg |